| Trace | Less than 1000 ppm CO2 in the background gas | 311, 316, 317, 318, 3160 | B-2C |
| 3000T series | L-2C | ||
| More than 1000 ppm up to 99% CO2 in the background gas | All trace oxygen instruments | A-2C | |
| Note: To use the A-2C or B-2C sensor in a 3000Tseries instrument, adapter part number B66378 is required. | |||
| Note: The L-2C sensor cannot be used in our olderanalog models (300 series). | |||
| Percent | Less than 1000 ppm CO2 in the background gas | 311P, 320A / P / BRC, 326, 327, 328 | B-1, B-3 standard; C-3 for high concentration of O2 |
| More than 1000 ppm up to 99% CO2 in the background gas | A-5 | ||
| CLASS (click for spec control drwg.) | PART No. | APPLICATION | COMMENTS |
| A-1 oxygen sensor | C6689-A1 | Ultra-fast response (90% in less than 4 seconds)in the percent oxygen region. | Use only when B-1 response time is insufficient. |
| A-2C oxygen sensor | C6689-A2C | Used for trace O2 measurements in CO2 containingbackgrounds (0.1-99% CO2). CO2 must be present. | Use where acid gases are present (CO2, HCl, HCN,etc). CO2 must be present. |
| A-2CXL oxygen sensor | B74033-A2CXL | Specially qualified A2C sensor to operate in XLmodel analyzers. Low range limit is 0-1 ppm. | Use where acid gases are present (CO2, HCl, HCN,etc). CO2 must be present. |
| A-2Cl:Obsolete, use A-2C | |||
| A-3:Obsolete, use A-5 | |||
| A-5 oxygen sensor | C6689-A5 | Intermittent or continuous monitoring of percentO2 in gas mixtures containing 0-99% CO2. 90% response is time is45 seconds. | Has more tolerance to exposure to air than A-3. Norestrictions on CO2 concentration. |
| B-1 oxygen sensor | C6689-B1 | General purpose percent oxygen analysis where fastresponse is desired (90% in less than 7 seconds). | Use B-3 or C-3 where longer life is required andresponse time is relatively unimportant. May be used to spot checkflue gas. |
| B-2C oxygen sensor | C6689-B2C | General purpose trace O2 analysis for use in inertgas and hydrocarbon streams. Low range limit is 0-10 ppm. | May be used in CO2 free gas streams containinghighly mobile gases such as H2, He and ethylene. |
| INSTA TRACE oxygen sensor | B71875 | Special patented construction provides rapidrecovery, during initial cell installation, to low PPM levels byreducing the recovery period from several hours to just minutes.Low range limit is 0-10 ppm. | Used for trace O2 analysis in inert gas andhydrocarbon streams. |
| B-2CXL oxygen sensor | C6689-B2CXL | Specially qualified B-2C sensor to operate in XLmodel analyzers. Low range limit is 0-1 ppm. | May be used in CO2 free gas streams containinghighly mobile gases such as H2, He and ethylene. |
| B-3 oxygen sensor | C6689-B3 | Percent O2 analysis, best compromise between longlife and fast response. | May be used to spot check flue gas. |
| B-4C oxygen sensor | C14613-B4C | Trace O2 cell specifically designed for the Model317EXI. | Same performance as the B-2. |
| B-5 oxygen sensor | C45091-B5 | Special application cell (non-standard) where longlife and moderately fast response are required; B-5 has a factor of3x the life of a B-3. | Requires 3x the amplifier gain of the B-3.Expected life is 36 months in air. |
| B-5F oxygen sensor | C45091-B5F | Special application cell (non-standard) where longlife and moderately fast response, in the presence of CO2, arerequired. | Uses Type F (acid electrolyte). Requires higheramplifier gain than normal. 90% response time is 13 seconds.Expected life in air is 6 months. |
| B-7 oxygen sensor | B45091-B7 | Special application cell (non-standard) where fastresponse is required; the B-7 has a factor of 3x the life of a B-1cell. | Requires 3x the amplifier gain of the B-1.Expected life in air is 24 months. |
| B-7F oxygen sensor | B45091-B7F | Special application cell (non-standard) wheremoderate life and fast response, in the presence of CO2, arerequired. | Requires higher amplifier gain than normal. 90%response time is 7 seconds. Expected life in air is 3 months. |
| C-1 oxygen sensor | C6689-C1 | Rarely used. Schematic for referenceonly. | |
| C-2 oxygen sensor | C6689-C2 | Rarely used. Schematic for referenceonly. | |
| C-3 oxygen sensor | C6689-C3 | Percent O2 measurements where maximum cell life isdesired (high O2 applications). | 90% response time is 30 seconds. Expected life inair is 18 months. |
| C-5 oxygen sensor | C45091-C5 | Special application cell (non-standard) where longlife is required; the C-5 has a factor of 3x the life of a C-3cell. | Requires 3x the amplifier gain of the C-3.Expected life in air is 54 months. |
| C-5F oxygen sensor | C45091-C5F | Special application cell (non-standard) where long life, in thepresence of CO2, is required. | Requires higher amplifier gain than normal. 90% response timeis 30 seconds. Expected life in air is 9 months. |
| C-6 oxygen sensor | C54090-C6 | Motion insensitive percent O2 measurements for use in hand heldportable monitors. | Best cell for OEM safety applications. |
| E-1 oxygen sensor | C14991-E1 | Special cell with performance characteristic of the C-3 with 5xthe life. | Cell is 2 to 3 times the length of standard MFCs; requiresspecial mounting. Designed for medical oxygen concentrators. |
| E-2 oxygen sensor | C57283-E2 | Long life cell for use is the Model 3290 (for use in unattendedair liquefaction nitrogen skids). | Has a 3 to 4 years expected life in air. Minimum range 0-1000ppm. |
| F-1R | C40371-F1R | 0-1000 ppm CO in Air | Used in Max 5 and 9950 Portable Combustion Analyzers |
| F-3 | C55194-F3 | Rarely used. Schematic for referenceonly. | |
| INSTA TRACE - CO2 | B73016 | Special patented construction provides rapidrecovery, during initial cell installation, to low PPM levels byreducing the recovery period from several hours to just minutes.Used for trace O2 measurements in CO2 containing backgrounds(0-99% CO2) in the 3000T analyzer series. | Use where acid gases are present (CO2, HCl, HCN,etc). CO2 must be present. The INSTA TRACE Micro-fuel Cell oxygensensor is capable of measuring below 1 ppm of oxygen within 15minutes of installation. |
| INSTA TRACE - XL oxygen sensor | B73592 | Special patented construction provides rapidrecovery, during initial cell installation, to low PPM levels byreducing the recovery period from several hours to just minutes.Specially qualified sensor to operate in XL model analyzers with alow range limit of 0-1 ppm. | Used for trace O2 analysis in inert gas andhydrocarbon streams. |
| K-1 oxygen sensor | C43832-K1 | Smallest Micro-fuel Cell, for use in applications where smallsize or multiple sensing requirements exist. | OEM applications; PC board mountable. |
| K-2:Obsolete | |||
| L-2C oxygen sensor | C6689-L2C | Used for general purpose trace oxygen analysis in the 3000Tanalyzer series. | Has a 2 year expected life. |
| L-2CXL oxygen sensor | C6689-L2CXL | Specially qualified L2C sensor to be used in 3000XL analyzerseries. Low range limit is 0-1 ppm. | May be used in CO2 free gas streams containing highly mobilegases such as H2, He and ethylene. |
| L-2CI oxygen sensor | C6689-L2CI | Used in the Model 3000T series for trace O2 measurements; useonly in CO2 and H2 background mixtures (e.g. inert gas generators). | CO2 concentration should be within +/-20% about some nominalconcentration. |
| S-2C oxygen sensor | C6689-S2C | Trace oxygen cell for use in TurbO2 analyzer. Provides rapidrecovery from air calibration (approximately 20 minutes). | Low range limit is 0-100 ppm. Has three pin contact system. |
| Z-2C oxygen sensor | C6689-Z2C | Has the same features as the S-2, except configured for use inthe Model 3190. | Low range limit is 0-100 ppm. Has normal Micro-fuel Cellcontact plate. |